Ultraviolet-Visible Spectroscopy for Medical Devices: What It Is and How It Helps

Ultraviolet-visible spectroscopy, or UV-Vis, is one of the simplest and most reliable tools in chemistry and materials testing. It is quick to run, easy to understand, and widely used in regulated environments.

Medical device teams see UV-Vis show up often in chemical characterization reports, but the purpose can feel unclear at first. This guide breaks it down in practical terms so you know what the test shows and why it matters.

UV-Vis spectroscopy is one of the analytical tools commonly used in chemical characterization for medical devices, especially when teams need fast information about concentration, purity, color, or material changes.

What is UV-Vis Spectroscopy?

UV-Vis spectroscopy measures how much ultraviolet and visible light a sample absorbs or passes through at different wavelengths. It is a simple way to understand how a material interacts with light.

Light moves through a sample, certain wavelengths are absorbed by the molecules, and the detector measures what gets through. The resulting pattern reflects the identity or concentration of substances in the sample.

This method is often called UV-Vis spectrophotometry. It uses the Beer–Lambert law to relate absorbance to concentration, which makes quantitative measurements possible.

Key concepts:

  • Wavelength range is typically 190 to 800 nanometers
  • Absorbance and transmittance show how much light is taken in or passed through
  • Spectra appear as curves that reflect how the sample responds to light

Why UV-Vis Matters in Medical Device Programs

UV-Vis is widely used in medical device testing because it is fast, relatively low-cost, and supported by long-standing pharmacopeial practices. These foundations make it a trusted, routine tool for regulated environments.

The method helps reveal key features of materials and solutions. It can show:

  • concentration
  • purity
  • color
  • clarity
  • early signs of degradation

These signals are often used to monitor device materials, coatings, extractables, and leachables throughout development.

UV-Vis also supports chemical characterization under ISO 10993-18. Teams often include it in extractables and leachables work or in performance and stability studies where color or concentration changes matter.

Typical goals include:

  • Confirming that a specific leachable or extractable stays below the intended concentration
  • Checking that coating or dye lots remain consistent in color and absorbance
  • Tracking oxidation or degradation of materials by watching for spectral shifts over time

The Core UV-Vis Tests You Will Hear About

UV-Vis is used in several practical ways during device development and chemical characterization. These are the tests that show up most often in reports and study plans.

Simple Absorbance and Concentration Measurements

This is the classic use of UV-Vis. The instrument measures absorbance at a specific wavelength. The result is compared to a calibration curve or interpreted through the Beer–Lambert law to determine concentration. It is direct, fast, and reliable.

You will see this approach used in several common situations:

  • Extractables and leachables solutions
  • Active ingredient or additive levels in coatings
  • Residuals in cleaning or reprocessing fluids

These applications rely on clear changes in absorbance that link concentration to how the material interacts with light at a chosen wavelength.

Full Spectrum Scans

A full-spectrum scan records absorbance across a wide range of wavelengths rather than at a single point. This creates a curve that shows how the sample interacts with light from the ultraviolet range into the visible range.

Full scans are helpful when you need a broader look at the material. They are often used to:

  • Check for unexpected peaks that may indicate extra components
  • Compare different material lots to confirm consistency
  • Follow changes that occur during aging, light exposure, or sterilization

This type of scan gives teams a fuller picture of how a sample behaves and helps spot changes that a single-wavelength measurement might miss.

Color and Clarity Assessments

UV-Vis is also used to measure color and clarity in transparent or translucent materials and solutions. The instrument detects small changes in absorbance that relate to turbidity or shifts in color, which makes it useful for quality checks and stability work.

Common applications include:

  • Tracking color stability in device housings or tubing
  • Measuring the clarity of solutions or rinses
  • Monitoring discoloration that can occur after UV or gamma sterilization

These assessments help teams understand how materials behave over time and whether a process or sterilization method affects appearance or transparency.

Time-Based or Kinetic Measurements

Some UV-Vis tests measure how a sample changes over time. The instrument takes repeated readings and tracks how absorbance shifts as a reaction or process moves forward. This approach is useful when you need to follow a change rather than capture a single snapshot.

Examples include:

  • Watching the degradation of an active coating in solution
  • Measuring the release rate of a drug from a device in model fluids

Kinetic measurements help teams understand how fast a change happens and whether that change stays within acceptable limits for performance or safety.

How UV-Vis Testing Is Actually Run (High-Level View)

Step 1. Sample and Blank Preparation

The process begins with preparing the sample in a suitable solvent or extraction medium. The sample needs to be well mixed and free of particles that could scatter light.

A blank is also prepared. The blank contains only the solvent or extraction medium. It serves as the baseline for the measurement, so the instrument can see how much of the light change comes from the sample itself.

Step 2. Instrument Setup

After the samples and blanks are ready, the instrument settings are chosen. This includes selecting the wavelength or the full scan range, depending on what the test needs to measure.

The path length and cuvette material are also chosen at this stage. Plastic, glass, or quartz cuvettes may be used. The choice depends on the wavelength and the type of sample being tested.

Step 3. Measurement and Data Capture

The blank is measured first to create a baseline. This tells the instrument how the solvent or extraction medium behaves with light on its own.

Next, the sample is measured. The instrument records the absorbance at the chosen wavelength or collects the full spectrum across the selected range. These values form the basis for all later interpretation.

Step 4. Interpretation

Once the data is collected, the absorbance can be converted to concentration if a calibration curve is available. This is one of the most common uses of UV-Vis because it provides quick, quantitative information.

Spectra can also be compared across lots or time points. Differences in peak height, peak position, or overall shape can show changes in materials or solutions.

The final step is deciding whether the results meet the predefined criteria for the study. These principles are central to many UV-Vis spectrophotometry fundamentals and applications used in analytical and materials labs.

Where UV-Vis Fits in Medical Device Testing Workflows

Chemical Characterization and Extractables / Leachables

UV-Vis often appears alongside other analytical techniques such as GC-MS, LC-MS, ICP-MS, and FT-IR when evaluating chemicals that may be released from medical devices under ISO 10993-18. It provides fast, targeted information, especially when a leachable has a clear absorbance peak that makes quantification straightforward.

It is especially useful when a leachable has a clear absorbance peak. In these cases, UV-Vis can give quick, reliable concentration information without needing a more complex method.

Supporting Biocompatibility and Hemocompatibility Tests

UV-Vis is often used behind the scenes in biocompatibility work. A common example is hemolysis testing, where the amount of released hemoglobin is measured spectrophotometrically to show how a device interacts with blood.

This type of analysis connects basic analytical chemistry to biological endpoints and helps confirm that a device meets safety expectations. UV-Vis plays a role in hemolysis testing, even when the technique is not the main focus of the study.

Cleaning and Reprocessing Validation

UV-Vis can help quantify cleaning agent residues, dyes, or other markers in rinse water during automated cleaning validation. It is a fast way to confirm that the cleaning process removes the chemicals or colorants used during manufacturing or reprocessing.

Strengths and Limitations Compared to Other Techniques

Strengths

  • Fast and relatively low in cost
  • Simple sample preparation
  • Non-destructive for many materials
  • Good for routine quantification when the analyte has a clear chromophore
  • Easy to integrate into extractables and leachables workflows

Limitations

  • Provides limited structural information compared to MS or NMR
  • Overlapping spectra can make interpretation difficult
  • Matrix components may interfere with absorbance signals
  • Requires calibration or standards for accurate quantitative work

Getting Good UV-Vis Data: Practical Tips for Device Teams

A few small habits can make UV-Vis results much more reliable.

  • Use clean cuvettes and check them for scratches before each run
  • Match blanks and samples in the same solvent and under the same extraction conditions
  • Avoid highly scattering samples unless the method is designed to handle them
  • Confirm the linearity range so that high concentrations do not produce misleading absorbance values

These steps help keep the data stable, consistent, and easier to interpret across lots and time points.

Common Missteps in UV-Vis Testing (And How to Avoid Them)

Even simple methods can produce bad data if a few basics are missed. These are the issues that show up most often.

  • Choosing the wrong wavelength for the analyte
  • Using dirty, scratched, or mismatched cuvettes
  • Ignoring baseline drift or skipping the blank measurement
  • Over-interpreting noisy spectra or very small peaks
  • Treating UV-Vis as the only answer when other techniques may be better suited

Avoiding these problems helps ensure that the data supports clear decisions rather than creating confusion.

FAQs About UV-Vis for Medical Device Teams

What is ultraviolet-visible spectroscopy used for?

UV-Vis is used to measure how a material absorbs or transmits light. This helps determine concentration, purity, color, clarity, and changes in materials over time. In medical devices, it often supports chemical characterization, extractables and leachables studies, and stability work.

What types of samples can UV-Vis analyze?

UV-Vis can analyze liquids, solutions, and transparent or translucent materials. It works well for extracts, leachables, cleaning fluids, coatings, dyes, and some polymer solutions. The sample should allow light to pass through without heavy scattering.

What is the difference between UV spectroscopy and UV-Vis spectroscopy?

UV spectroscopy focuses only on ultraviolet wavelengths. UV-Vis covers both ultraviolet and visible light. The wider range allows UV-Vis to capture more information, including color changes and absorbance peaks that occur in the visible region.

How do you perform UV-Visible spectroscopy?

A sample is prepared in a suitable solvent and placed in a cuvette along with a matching blank. The instrument measures absorbance at a selected wavelength or across a full spectrum. The data is then interpreted to determine concentration or to compare material behavior across samples.

When is UV-Vis the right tool, and when is another method better?

UV-Vis is best when the analyte has a clear absorbance peak and the goal is simple quantification or material comparison. More complex questions, such as identifying unknown compounds, usually require GC-MS, LC-MS, or FT-IR.

Does UV-Vis replace other analytical methods?

Not usually. UV-Vis works well for quick checks, but techniques like MS or FT-IR provide deeper structural details. Many extractables and leachables programs use UV-Vis alongside other tools.

How many samples do I need for meaningful UV-Vis testing?

Most studies use at least triplicate measurements and multiple lots when comparing materials. The exact number depends on the study design and the level of confidence needed.

What does UV-Vis tell me about material changes?

Changes in color, absorbance peaks, or spectrum shape can signal oxidation, degradation, or lot-to-lot shifts. UV-Vis is often used to confirm that materials remain stable after sterilization or aging.

Using UV-Vis as a Decision Tool

UV-Vis may seem like a basic analytical method, but it can offer valuable insight when used with purpose. It helps confirm consistency in materials, reveals small changes that affect device performance, and supports chemical characterization under ISO 10993-18. When teams understand what the test shows, the data becomes easier to interpret and much more useful.

A thoughtful approach makes UV-Vis more than a routine step. With the right wavelength choices, clean sample prep, and careful comparisons across lots or time points, it becomes a reliable decision tool that supports safety, stability, and regulatory confidence.

 

North American Biomedical Institute
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.